Nov 06, 2023
395
- Purpose
- This work instruction describes the process for altering the RCF settings in the MX5 centrifuge.
- This work instruction describes the process for altering the RCF settings in the MX5 centrifuge.
- Equipment/Software
- None.
- None.
- Instructions
- Press the Stop and Enter buttons at the same time
- Press the Up button until the speed shows 00008
- Press the RCF/RPM button to move to the next digit
- Press the Up button until the speed shows 00018
- Press the Enter button
- Press the Set button – P31 will blink
- Press the Up button to change P31 to P33 - P33 is rotor 1 RCF calculation
- Press the Set button
- Press the Up or Down button to change the flashing value. Press the RCF/RPM button to move from digit to digit. Final value should be 1510
- Press Set button
- Press the Up button to change from P33 to P35 - P35 is rotor 2 RCF calculation
- Press the Set button
- Press the Up or Down button to change the flashing value. Press the RCF/RPM button to move from digit to digit. Final value should be 1170
- Press Set button
- Press the Up button to change P35 to P37 - P37 is rotor 3 RCF calculation
- Press the Set button
- Press the Up or Down button to change the flashing value. Press the RCF/RPM button to move from digit to digit. Final value should be 1460
- Press Set button
- Press the Up button to change from P37 to P39 - P39 is rotor 4 RCF calculation
- Press Set button
- Press the Up or Down bu9on to change the flashing value. Press the RCF/RPM button to move from digit to digit. Final value should be 1250
- Press Set button 23. Press Enter button
- Additional Information
- Rotor 1 is identified as P33
- Rotor 2 is identified as P35
- Rotor 3 is identified as P37
- Rotor 4 is identified as P39
- Rotor 5 is identified as P41
- Rotor Radius in units of 0.1mm
SNR-7.5.1-W-019 | Rev 0