MXU Motor Drive Board Repair Instructions - SNR-7.5.1-W-005
print icon

 

Purpose

This work instruction describes the process for replacing the MXU Fuse. These repair instructions are valid for revision C MXU Digital Centrifuges.

Required Equipment

  • Fuse; 3A; 250V (Part # 100084)
  • 1/8” Flathead Screwdriver

Summary of Process

The procedure involves unplugging the unit, removing the motor control board, disconnecting and replacing the fuse, and then reassembling and testing the centrifuge.

 