1. Purpose 1. This work instruction describes the process for replacing the MXU Motor Drive Board. 2. These repair instructions are valid for revision C MXU Digital Centrifuges. For revisions older than C, please consult Product Development. 2. Equipment/Software 1. ITEM # PART # DESCRIPTION QTY. …
1. Purpose 1. This work instruction describes the process for replacing the lid latch on MX12 centrifuge units. 2. Equipment/Software 1. Replacement Lid Latch (101078-D) 2. Attachment Screws, M4 x 8mm, button head hex drive, stainless steel 3. 2.5mm hex driver 3. Instructions 4. Check that…
When it comes to the Turbo Mixer and the issue of running too fast or too slow, the issue is almost always attributed to the Potentiometer; this can be found at the bottom of the unit and can be accessed with a small Philips head screwdriver. 1. Ensure that the unit is powered on and the speed knob is turned all th…
1. Purpose 1. This work instruction describes the process for converting the base of MX12 centrifuge units. 2. These repair instructions are valid for Revision G MX12 Centrifuges. For revisions older than G, please consult Product Development. 2. Equipment/Software: 1. ITEM # PART # DESCRIPTION QT…