Home > Centrifuges > E8 Centrifuges > E8F and E8V Motor Drive Board Repair Instructions - SNR-7.5.1-W-008
E8F and E8V Motor Drive Board Repair Instructions - SNR-7.5.1-W-008
print icon

 

Purpose

This work instruction describes the process for replacing the soft start motor drive board in the E8 Fixed Centrifuge and E8 Variable Centrifuge. These instructions are valid for Revision J of the E8 Fixed and E8 Variable models.

 

Required Equipment

  • Motor Drive Board (Part # 101283)
  • #1 Philips Screwdriver

Summary of Process

The repair procedure involves unplugging the centrifuge, removing the rotor and base assembly, and then disconnecting the old motor drive board. A new motor drive board is installed, rewired, and the unit is reassembled. The process concludes with verifying that all centrifuge functionalities are working correctly.

 