Purpose This work instruction describes the process to replace the door magnet of 10L, 20L, and 30L Incubators. Required Equipment * Stainless Steel Covered Magnet (Part # 101407) * 2 mm Hex Key Summary of Process The procedure involves opening the incubator door, unscrewing the old magnet with a 2mm Hex Key,…
Purpose This work instruction describes how to replace a power switch for all versions of the E8 centrifuge. Required Equipment * Switch; power input; round body; illuminated red (Part # 100099) * #1 Philips Screwdriver Summary of Process The procedure involves unplugging the unit, removing the rotor and ba…
Purpose This work instruction describes the process of replacing the MX5 lid latch. Required Equipment * The full list of required equipment is not available in the provided document content. Summary of Process The repair procedure involves replacing the lid latch on an MX5 Digital Centrifuge. The attached doc…
Purpose This work instruction describes the process for replacing the Motor Control Board on the MX5 Digital Centrifuge. The instructions apply to boards with part numbers 100918-B for 110V units and 101072 for 220V units. Required Equipment * Motor Control Board (Part # 100918-B for 110V or 101072 for 220V) Su…
Purpose This work instruction describes the process for replacing the label of the MX5 Centrifuge. These repair instructions are valid for Revision G of the MX5 Centrifuge. Summary of Process This document provides instructions for replacing the label on an MX5 Centrifuge. The attached document provides a detail…
Purpose This work instruction describes how to replace a speed/tach sensor on the E8V Revision J and E8F Revision J centrifuge models. Required Equipment * IR Sensor (Part # 101070) * #1 Phillips screwdriver Summary of Process The repair procedure involves unplugging the unit and removing the rotor and ba…
Purpose This work instruction describes the process of replacing the Power Input on the MX12 centrifuges. Required Equipment * The full list of required equipment is not available in the provided document content. Summary of Process This document provides instructions for replacing the power input on an MX…
Purpose This work instruction describes the process for repairing the Lid Switch of the E8 Fixed and the E8 Variable centrifuges. These instructions are valid for Revision K of both centrifuge models. Required Equipment * Lid Safety Switch (Part # 101281) * Lid Switch Bracket (Part # 101282-B) * RHBS; 18-8;…
Purpose This work instruction describes the process for replacing the User Interface Board (100919-C) of the MX5 Centrifuge. Required Equipment * User Interface Board (Part # 100919-C) Summary of Process This document provides instructions for replacing the user interface board of the MX5 Digital Centrifuge. T…
Purpose This work instruction describes the process for replacing the MXU Power Jack. These repair instructions are valid for revision C MXU Digital Centrifuges. Required Equipment * Power Jack Wiring Assembly (Part # 100300) * Phillips Head Screwdriver * Crescent Wrench Summary of Process The repair procedure…