E8T Motor Drive Board Repair Instructions - SNR-7.5.1-W-006
print icon

 

Purpose

This document provides instructions for replacing the soft start motor drive board in the E8 Touch Centrifuge. These instructions are valid for E8 Touch Revisions A and B.


Required Equipment

  • Motor Drive Board (Part # 101283)
  • #1 Philips Screwdriver


Summary of Process

The procedure involves unplugging the unit, removing the rotor and base/motor assembly, and then disconnecting and replacing the motor drive board. The new board is rewired and the unit is reassembled. The repair concludes with a functionality check of the centrifuge.

 