Power Input Replacement Instructions - SNR-7.5.1-W-035 MX12
Purpose This work instruction describes the process of replacing the Power Input on the MX12 centrifuges. Required Equipment * The full list of required equipment is not available in the provided document content. Summary of Process This document provides instructions for replacing the power input on an MX…
-
E8D Motor Drive Board Repair Instructions - SNR-7.5.1-W-010
Purpose This work instruction describes the process for replacing the soft start motor drive board in the E8 Digital Centrifuge. These instructions are valid for Revision J of the E8 Digital model. Required Equipment * Motor Drive Board (Part # 100480-C) * #1 Philips Screwdriver Summary of Process The repai…
-
E8F & E8V Timer Repair Instructions - SNR-7.5.1-W-016
Purpose This work instruction describes the process for repairing the Timer Knob of the E8 Fixed and E8 Variable Centrifuges. These instructions are valid for Revision J of both models. Required Equipment * Timer (Part # 100203-B) * 10 mm socket * #1 Philips screwdriver Summary of Process The repair proced…
-
E8T Motor Drive Board Repair Instructions - SNR-7.5.1-W-006
Purpose This document provides instructions for replacing the soft start motor drive board in the E8 Touch Centrifuge. These instructions are valid for E8 Touch Revisions A and B. Required Equipment * Motor Drive Board (Part # 101283) * #1 Philips Screwdriver Summary of Process The procedure involves unpluggi…
-
E8F and E8V Lid Switch Repair Instructions - SNR-7.5.1-W-017
Purpose This work instruction describes the process for repairing the Lid Switch of the E8 Fixed and the E8 Variable centrifuges. These instructions are valid for Revision K of both centrifuge models. Required Equipment * Lid Safety Switch (Part # 101281) * Lid Switch Bracket (Part # 101282-B) * RHBS; 18-8;…
-
Ready Warm Power Jack - SNR-7.5.1-W-046
Purpose This document provides instructions for replacing the power jack for the Ready Warm Tissue Bath. These instructions are valid for revision 6 of the Ready Warm. Required Equipment * The full list of required equipment is not available in the provided document content. Summary of Process This work instru…
-
Incubator Magnet Replacement Instructions - SNR-7.5.1-W-026
Purpose This work instruction describes the process to replace the door magnet of 10L, 20L, and 30L Incubators. Required Equipment * Stainless Steel Covered Magnet (Part # 101407) * 2 mm Hex Key Summary of Process The procedure involves opening the incubator door, unscrewing the old magnet with a 2mm Hex Key,…
-
MXU Power Jack Repair Instructions - SNR-7.5.1-W-014
Purpose This work instruction describes the process for replacing the MXU Power Jack. These repair instructions are valid for revision C MXU Digital Centrifuges. Required Equipment * Power Jack Wiring Assembly (Part # 100300) * Phillips Head Screwdriver * Crescent Wrench Summary of Process The repair procedure…
-
MXU Speed Sensor Repair Instructions - SNR-7.5.1-W-012
Purpose This work instruction describes the process for replacing the MXU Speed Sensor on a revision C MXU Digital Centrifuge. Required Equipment * Speed Sensor (Part # 100271-C) * Phillips Head Screwdriver Summary of Process The repair process involves opening the centrifuge, removing the rotor and base plate,…
-
MXU Speed Wheel Repair Instructions - SNR-7.5.1-W-013
Purpose This document provides instructions for replacing the MXU Speed Wheel on revision C MXU Digital Centrifuges. Required Equipment * Speed Sensor Wheel (Part # 101103) * Speed Sensor Hub (Part # 101104) * Speed Sensor Hub Stop (Part # 101105) * Blind Rivet, SS (Part # 100787) * Phillips Head Screwdri…