Home > Centrifuges > E8 Centrifuges > E8F & E8V Timer Repair Instructions - SNR-7.5.1-W-016
E8F & E8V Timer Repair Instructions - SNR-7.5.1-W-016
print icon

 

Purpose

This work instruction describes the process for repairing the Timer Knob of the E8 Fixed and E8 Variable Centrifuges. These instructions are valid for Revision J of both models.

 

Required Equipment

  • Timer (Part # 100203-B)
  • 10 mm socket
  • #1 Philips screwdriver
     

Summary of Process

The repair procedure involves ensuring the unit is unplugged and unpowered , removing the rotor and base/motor assembly , disconnecting and replacing the old timer , and then reassembling the centrifuge. The repair concludes with a verification of all functionalities under normal laboratory conditions.

 