Purpose This work instruction describes the process for replacing the soft start motor drive board in the E8 Fixed Centrifuge and E8 Variable Centrifuge. These instructions are valid for Revision J of the E8 Fixed and E8 Variable models. Required Equipment * Motor Drive Board (Part # 101283) * #1 Philips Scr…
Purpose This work instruction describes the process for replacing the soft start motor drive board in the E8 Digital Centrifuge. These instructions are valid for Revision J of the E8 Digital model. Required Equipment * Motor Drive Board (Part # 100480-C) * #1 Philips Screwdriver Summary of Process The repai…
Purpose This work instruction describes the process for replacing the lid of all E8 Centrifuges. Required Equipment * The full list of required equipment is not available in the provided document content. Summary of Process The procedure involves the complete process for replacing the lid on any E8 Centrifug…
Purpose This work instruction describes the process for replacing the lid gasket of MX12 centrifuge units. Required Equipment * Lid Gasket (Part # from content) Summary of Process The procedure involves removing the old lid gasket and installing a new one to ensure proper function and seal of the centrifuge …
Purpose This work instruction describes the process of replacing the Power Switch (100099) on the Slide Warmer. Required Equipment * Power Switch (Part # 100099) Summary of Process This document provides instructions for replacing the power switch on a Slide Warmer. The attached document contains the detailed …
Purpose This work instruction describes the process for replacing the MXU Motor Control Board. These repair instructions are valid for revision C MXU Digital Centrifuges. Required Equipment * Motor Control Board (Part # 100484) * Needle Nose Pliers * Screwdriver Summary of Process The repair procedure invo…
Purpose This work instruction describes the process of replacing the Power Switch for both the CytoPrep-5 and CytoPrep-10 models. Required Equipment * Power Switch (Part # 100099) Summary of Process This document provides instructions for replacing the power switch on both the CytoPrep-5 and CytoPrep-10 units.…
Purpose This work instruction describes the process for repairing the Timer Knob of the E8 Fixed and E8 Variable Centrifuges. These instructions are valid for Revision J of both models. Required Equipment * Timer (Part # 100203-B) * 10 mm socket * #1 Philips screwdriver Summary of Process The repair proced…
Purpose This work instruction describes the process of replacing the Fuse on the MX12 centrifuges. Required Equipment * The full list of required equipment is not available in the provided document content. Summary of Process This document provides instructions for replacing the fuse on an MX12 centrifuge. T…
Purpose This work instruction describes the process to replace the User Interface Board on the MX12 Centrifuges. Required Equipment * User Interface Board (Part # 100919) Summary of Process This document provides instructions for replacing the user interface board. The procedure involves opening the unit, disc…