Purpose This document describes the procedure for replacing the E8 Lid Latch. These instructions are valid for E8V rev F, E8F rev D, and E8T rev F. Required Equipment * Latch (Part # 100073) * Cam Latch; 1.5” (Part # 100473) * #1 Screwdriver * Loc-Tite Blue 425 * Hios CL-6500 electric driver with #3 Phill…
Purpose This work instruction describes the process to replace a power jack for all versions of the E8 Centrifuge. Required Equipment * Power Jack Cable Assembly; E8 (Part # 100525) * #1 Philips Screwdriver Summary of Process The procedure involves unplugging the unit, removing the rotor and base/motor asse…
Purpose This work instruction describes the process for replacing the hinge on MX12 centrifuge units. Required Equipment * Black Plastic Hinge; adjustable friction (Part # 101323) * Machine Screw; M4 X 0.7; 8mm; Black Oxide (Part # 100996) * 2.5mm hex driver Summary of Process The procedure involves turning …
Purpose This work instruction describes the process for replacing the label of the MX5 Centrifuge. These repair instructions are valid for Revision G of the MX5 Centrifuge. Summary of Process This document provides instructions for replacing the label on an MX5 Centrifuge. The attached document provides a detail…
Purpose This work instruction describes the process of replacing the Fuse on the MX12 centrifuges. Required Equipment * The full list of required equipment is not available in the provided document content. Summary of Process This document provides instructions for replacing the fuse on an MX12 centrifuge. T…
Purpose This work instruction describes the process of replacing the motor in the MX12 Centrifuge. Required Equipment * Motor (Part # 100968-B or 100994-B) Summary of Process The procedure involves replacing the motor in an MX12 centrifuge. The attached document provides a step-by-step guide on how to perform …
Purpose This work instruction describes the process for replacing the soft start motor drive board in the E8 Digital Centrifuge. These instructions are valid for Revision J of the E8 Digital model. Required Equipment * Motor Drive Board (Part # 100480-C) * #1 Philips Screwdriver Summary of Process The repai…
Purpose This work instruction describes the process for replacing the MXU Speed Sensor on a revision C MXU Digital Centrifuge. Required Equipment * Speed Sensor (Part # 100271-C) * Phillips Head Screwdriver Summary of Process The repair process involves opening the centrifuge, removing the rotor and base plate,…
Purpose This work instruction describes the process for repairing the Timer Knob of the E8 Fixed and E8 Variable Centrifuges. These instructions are valid for Revision J of both models. Required Equipment * Timer (Part # 100203-B) * 10 mm socket * #1 Philips screwdriver Summary of Process The repair proced…
Purpose This work instruction describes the process to replace the door magnet of 10L, 20L, and 30L Incubators. Required Equipment * Stainless Steel Covered Magnet (Part # 101407) * 2 mm Hex Key Summary of Process The procedure involves opening the incubator door, unscrewing the old magnet with a 2mm Hex Key,…