You are using an unsupported browser. Please update your browser to the latest version on or before July 31, 2020.
close
You are viewing the article in preview mode. It is not live at the moment.
Welcome to the LW Scientific Support Portal

Showing articles from centrifuge tag

MX5 Motor Repair Instructions - SNR-7.5.1-W-045

Purpose This work instruction describes the process of replacing the motor in an MX5 Centrifuge. Required Equipment * The full list of required equipment is not available in the provided document content. Summary of Process This document provides instructions for replacing the motor on an MX5 Centrifuge. The a…

MX5 User Interface Board Repair Instructions - SNR-7.5.1-W-041

Purpose This work instruction describes the process for replacing the User Interface Board (100919-C) of the MX5 Centrifuge. Required Equipment * User Interface Board (Part # 100919-C) Summary of Process This document provides instructions for replacing the user interface board of the MX5 Digital Centrifuge. T…

E8 Digital Fuse Replacement Instructions - SNR-7.5.1-W-039

Purpose This work instruction describes the process for replacing the motor drive board fuse in the E8 Digital Centrifuge. These repair instructions are valid for the E8 Digital Centrifuge. Required Equipment * Fuse; 2A, 250V (Part # 100084) * 1/8” Flathead Screwdriver * #1 Philips Screwdriver Summary of Pr…

E8F and E8V Motor Drive Board Repair Instructions - SNR-7.5.1-W-008

Purpose This work instruction describes the process for replacing the soft start motor drive board in the E8 Fixed Centrifuge and E8 Variable Centrifuge. These instructions are valid for Revision J of the E8 Fixed and E8 Variable models. Required Equipment * Motor Drive Board (Part # 101283) * #1 Philips Scr…

E8F and E8V Lid Switch Repair Instructions - SNR-7.5.1-W-017

Purpose This work instruction describes the process for repairing the Lid Switch of the E8 Fixed and the E8 Variable centrifuges. These instructions are valid for Revision K of both centrifuge models. Required Equipment * Lid Safety Switch (Part # 101281) * Lid Switch Bracket (Part # 101282-B) * RHBS; 18-8;…

MXU Motor Control Board Repair Instructions - SNR-7.5.1-W-011

Purpose This work instruction describes the process for replacing the MXU Motor Control Board. These repair instructions are valid for revision C MXU Digital Centrifuges. Required Equipment * Motor Control Board (Part # 100484) * Needle Nose Pliers * Screwdriver Summary of Process The repair procedure invo…

MX12 Motor Replacement Instruction - SNR-7.5.1-W-034

Purpose This work instruction describes the process of replacing the motor in the MX12 Centrifuge. Required Equipment * Motor (Part # 100968-B or 100994-B) Summary of Process The procedure involves replacing the motor in an MX12 centrifuge. The attached document provides a step-by-step guide on how to perform …

E8D Motor Drive Board Repair Instructions - SNR-7.5.1-W-010

Purpose This work instruction describes the process for replacing the soft start motor drive board in the E8 Digital Centrifuge. These instructions are valid for Revision J of the E8 Digital model. Required Equipment * Motor Drive Board (Part # 100480-C) * #1 Philips Screwdriver Summary of Process The repai…

MXU Power Switch Repair Instruction - SNR-7.5.1-W-009

Purpose This work instruction describes the process for replacing the power switch (100413) on a revision C MXU Digital Centrifuge. Required Equipment * Power Switch (Part # 100413) * Phillips Head Screwdriver Summary of Process The repair procedure involves removing the rotor and tube shields, inverting the …

MX12 Lid Gasket Instruction - SNR-7.5.1-W-030

Purpose This work instruction describes the process for replacing the lid gasket of MX12 centrifuge units. Required Equipment * Lid Gasket (Part # from content) Summary of Process The procedure involves removing the old lid gasket and installing a new one to ensure proper function and seal of the centrifuge …