E8F and E8V Lid Switch Repair Instructions - SNR-7.5.1-W-017

Purpose This work instruction describes the process for repairing the Lid Switch of the E8 Fixed and the E8 Variable centrifuges. These instructions are valid for Revision K of both centrifuge models. Required Equipment * Lid Safety Switch (Part # 101281) * Lid Switch Bracket (Part # 101282-B) * RHBS; 18-8;…

Ready Warm Power Jack - SNR-7.5.1-W-046

Purpose This document provides instructions for replacing the power jack for the Ready Warm Tissue Bath. These instructions are valid for revision 6 of the Ready Warm. Required Equipment * The full list of required equipment is not available in the provided document content. Summary of Process This work instru…

E8 Lid Latch Replacement Instructions - SNR-7.5.1-W-027

Purpose This document describes the procedure for replacing the E8 Lid Latch. These instructions are valid for E8V rev F, E8F rev D, and E8T rev F. Required Equipment * Latch (Part # 100073) * Cam Latch; 1.5” (Part # 100473) * #1 Screwdriver * Loc-Tite Blue 425 * Hios CL-6500 electric driver with #3 Phill…

Motor Control Board - SNR-7.5.1-W-032

Purpose This work instruction describes the process to replace the Motor Control Board on the MX12 Centrifuges. Required Equipment * User Interface Board Summary of Process This document provides instructions for replacing the MX12 Motor Control Board. The procedure involves removing the existing board and ins…

E8 Digital Fuse Replacement Instructions - SNR-7.5.1-W-039

Purpose This work instruction describes the process for replacing the motor drive board fuse in the E8 Digital Centrifuge. These repair instructions are valid for the E8 Digital Centrifuge. Required Equipment * Fuse; 2A, 250V (Part # 100084) * 1/8” Flathead Screwdriver * #1 Philips Screwdriver Summary of Pr…

E8 Speed Wheel Replacement Instructions - SNR-7.5.1-W-028

Purpose This work instruction describes how to replace a speed wheel assembly, which includes two speed sensor hubs and one hub stop. Required Equipment * Speed Sensor Wheel (Part # 100482) * Speed Sensor Hub (Part # 100481) * Speed Sensor Hub Stop (Part # 100487) * #1 Philips Screwdriver Summary of Proc…

E8T Motor Drive Board Repair Instructions - SNR-7.5.1-W-006

Purpose This document provides instructions for replacing the soft start motor drive board in the E8 Touch Centrifuge. These instructions are valid for E8 Touch Revisions A and B. Required Equipment * Motor Drive Board (Part # 101283) * #1 Philips Screwdriver Summary of Process The procedure involves unpluggi…

MXU Fuse Repair Instruction - SNR-7.5.1-W-007

Purpose This work instruction describes the process for replacing the MXU Fuse. These repair instructions are valid for revision C MXU Digital Centrifuges. Required Equipment * Fuse; 3A; 250V (Part # 100084) * 1/8” Flathead Screwdriver Summary of Process The procedure involves unplugging the unit, removing the mot…

Service Light & Display Board Reset Instructions - SNR-7.5.1-W-002

A downloadable file is attached at the bottom of this article. It contains complete setup or service instructions, diagrams, and detailed part information. Overview This article explains how to reset the service light and display board on LW Scientific E8 Digital and Universal centrifuges. These reset instructions…

E8 Centrifuge Parts Diagrams

USA E8 Centrifuge Parts Diagrams If you’re looking for internal component diagrams or part numbers for your USA E8 centrifuge, you’re in the right place. This guide includes up-to-date diagrams for all current USA E8 models. You can use these diagrams to: * Identify part numbers * Understand how components are a…